Sample Requirements - Flow Cytometry

     ANALYSIS

             Negative controls
                    Cells without any added antibody (autofluorescence)
                    Cells plus a non-reactive antibody (non-specific binding control, also called isotype control).
             Positive control
                    Cells plus an antibody known to bind to the cells being used.
             Experimental
                    Cells plus antibody in question (including perhaps several dilutions)
             Other controls that are sometimes included
                    Cells that would not be expected to bind the experimental antibody being tested. For example, if
                    the antibody being tested is reactive with malignant cells, you may want to include a "normal" cell to
                    rule out non-specific binding.
 

     SORTING

           Flow Rate (cells/sec)  1,000     2,000     5,000     10,000

             Recovery (*)           95%       90%      70%        60%

             Purity (**)             99%        99%      99%        99%
 

     (*) the percentage of positive cells actually recovered during the sort.
     (**) the percentage of the recovered cells that are positive for the marker under investigation [CD4 in the example
     below].

     Example of sort recovery.
     A sample of 36 x 10E6 cells is brought to the FACS lab after being stained with a CD4 antibody; FACS analysis
     indicates the sample has 35% fluorescence-positive (CD4+) cells in it. The sample requires 1 hr of sorting time (at
     10,000 cells/sec). At 36% positivity, the theoretical yield of CD4 cells would be 12 x 10E6 cells: [(36 x 10E6) x
     (0.35)] At a recovery rate of 65%, we are able to collect 8 x 10E6 cells [(12 x 10E6) x (0.65)] which will be =/>99%
     pure, i.e., 99% of them will be CD4+.